Abstract

In a radiation individual monitoring program, the type testing of measuring devices is a great important part of the quality management system. The IEC-62387 standard applies to all dosimetry systems that measure external photon and/or beta radiation within limited ranges of the associated physical parameters. In this work, a type-testing program was conducted for a manual thermoluminescent dosimetry (TLD) reader employing the IEC-62387 performance criteria of radiation and environmental. The uncertainty of non-linearity of response of the dosimetry system in a range of 0.7-850 mSv was obtained between -15% and + 17% which fulfilled the IEC-62387 standard range of -16% to +18%. Furthermore, the total uncertainty of all reader tests was measured to be 12%, which was less than the criteria of 20% in the IEC-62387 standard. Thus, it can be concluded that the TLD reader met all requirements of the IEC standard for the reader-tests by an appropriate margin.

Keywords